Soft Contact Lens Measurement and Inspection Instruments
Sterling supplies a range of instruments for the accurate measurement and
inspection of soft contact lenses. All instruments use the best-practice method
of immersing the soft contact lens in a saline solution, using a wet cell for
lens analysis.
OPERATING PRINCIPLES OF OPTIMEC MACHINES
The lens diameter is measured and the surface is inspected as follows:
The soft contact lens is positioned on a 10mm diameter cylinder. A probe is
raised through the center of the cylinder via a micrometer screw and hand
operated dial (BCOR scale). The lens diameter is viewed on a projection screen.
The BCOR value is read from the BCOR scale.
A graticule (i.e. grid lines in the focal plane) is mounted within the wet cell
and is projected along with the plan view of the soft contact lens. This method
ensures accurate measurement of the lens diameter, irrespective of magnification
at the viewing screen. The smallest division is 0.1mm.
JCF and JCF/F
 JCF/F
(with Saline Filtration System) |
 JCF |
The JCF is available with or without a saline filtration system.
The JCF/F offers a peristaltic pump mounted on the cabinet, which feeds saline
back to the cell via a renewable cartridge filter. The pump system removes
particles from the saline, to ensure that the image of the contact lens in the
wet cell is not affected by these.
Both the JCF and JCF/F can be fitted with the TC20 temperature control system.
JCF and JCF/F Benefits:
- Easy to use wet cell system
- Measures all 3 critical parameters of soft contact lenses to +/- 0.05mm.
- Inside radius measurement 6.5mm to 9.5mm
- Center thickness measurement 3.5mm to maximum
- Overall diameter measurement 6.0mm to 16.0mm
- Plan view magnification X16
- Profile view magnification X19
- Dual projection system allows both lens profile and full edge inspection
- Upper and lower surface inspection by focusing through lens
- Optional saline filtration system
- Optional temperature control of wet cell
TC20: Temperature control for lens analysis
wet cells

TC20 (Temperature Control)
The TC20 is a consistent and accurate solid state-cooling
instrument, using the reliable Peltier effect. A temperature probe using an NTC
thermistor is immersed in the wet cell, and the saline is heated or cooled to
the required temperature. It is then maintained at that temperature at all
times.
The default temperature is 20 degrees Celcius, but the optional temperature
settings range from 15 to 38 degrees Celsius for special needs or Research and
Development applications.
Filters for these units are available from Sterling: contact us
Part number: OPT1001-1
TC20 Benefits:
-
Saline cooling for consistent 20 degree Celsius for normal
operation
-
Temperature constant to +/- 0.3 degrees Celsius
-
Avoids dimensional variations due to temperature change
-
Warming of saline in winter conditions
-
Heating or cooling of saline solution to required temperature
(15 to 38 degrees Celsius) for R&D and other applications
-
Temperature control unit is separate and can be removed/
switched on or off as required.
Chiltern

The Chiltern is an economical and cost-effective version of the
JCF, which operates on the same principles of measurements.
It cannot be fitted with saline filtration or temperature control, and the
design of the optics is simplified to reduce costs.
Nevertheless, it is an accurate and useful instrument for checking soft contact
lenses, and is used by both practitioners and manufacturers for a variety of
inspection tasks.
|